- HOME>
- PRODUCTS>
- Vibration Test Systems>
- EM-series(ECO-shaker)
Energy Saving Type Vibration Test Systems
EM-series(ECO-shaker)
- Control electricity consumption automatically depending on test conditions
- ECO-shaker realized full-scale energy-saving vibration test that was difficult in conventional shakers. It enables vibration test operated at automatically optimized electricity consumption by inputting usual test definition.

■Always developing vibration test systems
IMV's dynamic simulation systems have been developed with "intelligence" in mind and endeavour to support customers testing. The new Eco technology has been used in the IMV range.

Complicated settings are not needed.
- Electricity consumption might be managed manually on electrodynamic systems, but this could be quite difficult for the system operator to optimise the consumption due to payload and test conditions. Our ECO-Shaker can achieve optimisation of power consumption refering to payload conditions.
You can confirm the result of reduction of CO2 and electricity consumption at glance.
- By combination with IMV vibration controller ‘K2’ , ECO-shaker system will confirm real-time electricity consumption during testing. It was never available by conventional systems. Power consumption can be reported numerically after each test.
You can expect improvement of laboratory environment.
- Noise to be expected from the conventional test system will give limitaion to installation place selection. Our ECO-Shaker can suppress the noise level by controlling the engine blower at optimised speed. Optimum control of the cooling system will reduce power consumption and noise level refering to test conditions.
We will contribute to well being of society from both sides of quality and environmentt
- By Clean Development Mechanism (CDM) in January 2008 and revision of Rationalisation in Energy Use law in April 2009 were introduced in Japan, all enterprises have been obliged to be more energy efficient. ECO-Shaker promotes reduction of costs and CO? by saving electricity consumption, costs, and consequently contributes to society.
Real automatic energy saving driving
- Complicated settings are not needed.
- ECO-shaker is a electrodynamic vibration test system which automatically optimize output of power amplifier, power input to vibration and cooling blower rotation speed refering to payload and test conditions. So, complicated manual settings are not needed anymore. Since it is real automatic control according to working environment conditions, "drastic change in output levels due to change of test sample characteristics during the test", "temperature up in system caused by rising heat during operation" etc.. even if those phenomenon happen, it is possible to keep operation without stopping the test.

You can confirm reduction of CO2 and electricity consumption at real-time.
- You can confirm the result of reduction of CO2 and power consumption at glance.
- By combination with IMV vibration controller ‘K2’ , ECO-shaker system will confirm real-time electricity consumption during testing. It was never available by conventional systems. Power consumption can be reported numerically after each test, which clearly shown on monitor or display. So you can confirm the result at glance.

Figures are calculated by comparing system with conventional system of same sine force rating. Results may vary for systems, test conditions and cases.
- Comparison of electricity consumption with conventional shaker
- The less system output, the more saving-energy you can expect.


Operation of ISM-EM (Power consumption)
- Power consumption on conventional vibration systems would require manual calculation refering to test conditions. ISM-EM technology of the ECO-Shaker can automatically control the output of the power amplifier, force generation of the shaker, and rotation speed of the cooling blower to achieve best efficiency refering to test conditions.

Correspondence to existing system [Retro fit]
- Saving energy by ISM-EM technology can be applied on existing IMV vibration test systems by adding the ISM-EM unit and some calibrations. Contact us for further information and delivery.

ECO-Shaker lineup
- Specifications
| System Model | EM2201 | EM2301 | EM2401 |
EM2501 | EM2502 | EM2551 | EM2601 | ||
|---|---|---|---|---|---|---|---|---|---|
![]() |
Freq. Rang(Hz) | 0~3300 | 0~3000 | 0~2600 | 0~2500 | 0~2500 | 0~2600 5) | 0~2600 5) | |
![]() |
SINE(kN) | 8 | 16 | 24 | 32 | 40 | 49 | 54 | |
| RANDOM(kNrms)2) | 8 | 16 | 24 | 32 | 40 | 49 | 54 | ||
| SHOCK(kN) | 16 | 32 | 48 | 64 | 80 | 98 | 108 | ||
|
SINE(m/s2) | 1250 | 1250 | 1200 | 914 | 1142 | 907 | 1000 | |
| RANDOM(m/s2rms) | 875 | 875 | 840 | 640 | 800 | 635 | 700 | ||
| SHOCK (m/s2) | 2500 | 2500 | 2400 | 1828 | 2284 | 1814 | 2000 | ||
![]() |
SINE(m/s) | 2.2 | 2.2 | 2.2 | 2.2 | 2.2 | 2.2 | 2.2 | |
| SHOCK(m/s) | 2.2 | 2.2 | 2.2 | 2.2 | 2.2 | 2.2 | 2.2 | ||
![]() |
SINE(mmp-p) | 51 | 51 | 51 | 51 | 51 | 51 | 51 | |
| Max. TRAVEL(mmp-p) | 60 | 64 | 68 | 68 | 68 | 64 | 64 | ||
| Max. Load.(kg) | 200 | 300 | 400 | 600 | 600 | 1000 | 1000 | ||
| Power Requirements(kVA)※1 | 16.4 | 26 | 36 | 51 | 57 | 80 | 83 | ||
|
Model | i220 | i230 | i240 |
i250 | i250 | i255 | i260 | |
| Armature Mass(kg) | 6.4 | 12.8 | 20 | 35 | 35 | 54 | 54 | ||
| Armature Diameter(φmm) | 190 | 200 | 290 | 440 | 440 | 446 | 446 | ||
| Allowable eccentric moment(N・m) | 294 | 700 | 850 | 1550 | 1550 | 1550 | 1550 | ||
| Dimensions(mm)W×H×D | 1020× 903× 550 |
1124× 957× 860 |
1234× 997× 890 |
1463× 1187× 1100 |
1463× 1187× 1100 |
1527× 1196× 1100 |
1527× 1196× 1100 |
||
| Mass(kg) | 900 | 1500 | 2000 | 3000 | 3000 | 3500 | 3500 | ||
![]() |
Model | SA1M- i20EM |
SA2M- i30EM |
SA3M- i40EM |
SA4M- i50EM |
SA5M- i50EM |
SA7M- i55EM |
SA7M- i60EM |
|
| Max. Output(kVA) | 10 | 20 | 29 | 40 | 50 | 58 | 64 | ||
| Dimensions(mm)W×H×D | 580 1750 850 |
580 1750 850 |
580 1750 850 |
580 1950 850 |
580 1950 850 |
1160 1950 850 |
1160 1950 850 |
||
| Mass(kg) | 330 | 350 | 460 | 900 | 930 | 1400 | 1400 | ||
![]() |
Vibration Controller | See Vibration Controller K2 | |||||||
![]() |
Cooling Method | Air cooling | |||||||
![]() |
Dimensions(mm)W×H×D | 492 1128 627 |
606 1315 844 |
707 1531 896 |
1057 1841 1078 |
1057 1841 1078 |
1328 2140 1033 |
1328 2140 1033 |
|
| Mass(kg) | 70 | 125 | 210 | 250 | 250 | 370 | 370 | ||
- *1)Power supply required is 3-phase 200/220/240/380/400/415/440V,50/60Hz. Voltage Down Transformer (Step-down transformer) is required for other voltage.
- Rated force of RANDOM follow the ISO5344 standard. Please contact us for further information on RANDOM force conditions of each system.
- Each value of specification indicates maximum power of the system. In the case of a long hours of test, it needs to be operated less than 70% of maximum power.
- In the case of RANDOM durability test, please set the test definition of the peak value of acceleration waveform to be operated less than 1400 m/s2.
- In the band of more than 2000Hz, excitation force pitches at the rate of -6dB/oct.
(The continuous use could be the cause of a premature failure.)
★ E/L is required for exporting.(See P.4)
- Specifications (Large Displacement Type)
| System Model | EM2305 | EM2405 | EM2505 |
EM2506 | EM2555 | EM2605 | ||
|---|---|---|---|---|---|---|---|---|
![]() |
Freq. Rang(Hz) | 0~3000 | 0~2400 | 0~2200 | 0~2200 | 0~2600 5) | 0~2600 5) | |
![]() |
SINE(kN) | 16 | 24 | 35 | 40 | 49 | 54 | |
| RANDOM(kNrms)2) | 16 | 24 | 35 | 40 | 49 | 54 | ||
| SHOCK(kN) | 40 | 55 | 70 | 87 | 98 | 112 | ||
|
SINE(m/s2) | 941 | 923 | 777 | 888 | 777 | 857 | |
| RANDOM(m/s2rms) | 658 | 646 | 544 | 622 | 544 | 600 | ||
| SHOCK (m/s2) | 2352 | 2115 | 1555 | 1933 | 1555 | 1777 | ||
![]() |
SINE(m/s) | 2.4 | 2.4 | 2.4 | 2.4 | 2.4 | 2.4 | |
| SHOCK(m/s) | 2.4 | 2.4 | 2.4 | 2.4 | 2.4 | 2.4 | ||
![]() |
SINE(mmp-p) | 100 | 100 | 100 | 100 | 100 | 100 | |
| Max. TRAVEL(mmp-p) | 120 | 120 | 120 | 120 | 116 | 116 | ||
| Max. Load.(kg) | 300 | 400 | 600 | 600 | 1000 | 1000 | ||
| Power Requirements(kVA)※1 | 28 | 38 | 53 | 57 | 82 | 86 | ||
|
Model | J230 | J240 | J250 | J250 | i255 | i260 | |
| Armature Mass(kg) | 17 | 26 | 45 | 45 | 63 | 63 | ||
| Armature Diameter(φmm) | 200 | 290 | 440 | 440 | 446 | 446 | ||
| Allowable eccentric moment(N・m) | 700 | 850 | 1550 | 1550 | 1550 | 1550 | ||
| Dimensions(mm)W×H×D | 1124× 1079× 850 |
1234× 1145× 890 |
1463× 1301× 1100 |
1463× 1301× 1100 |
1527× 1319× 1100 |
1527× 1319× 1100 |
||
| Mass(kg) | 1800 | 2400 | 3500 | 3500 | 4100 | 4100 | ||
![]() |
Model | SA3M- J30EM |
SA4M- J40EM |
SA5M- J50EM |
SA6M- J50EM |
SA7M- J55EM |
SA7M- J60EM |
|
| Max. Output(kVA) | 23 | 34 | 50 | 57 | 64 | 70 | ||
| Dimensions(mm)W×H×D | 580 1750 850 |
580 1750 850 |
580 1950 850 |
580 1950 850 |
1160 1950 850 |
1160 1950 850 |
||
| Mass(kg) | 380 | 490 | 930 | 960 | 1400 | 1400 | ||
![]() |
Vibration Controller | See Vibration Controller K2 | ||||||
![]() |
Cooling Method | Air cooling | ||||||
![]() |
Dimensions(mm)W×H×D | 606 1315 844 |
707 1531 896 |
1057 1841 1078 |
1057 1841 1078 |
1328 2140 1033 |
1328 2140 1033 |
|
| Mass(kg) | 125 | 210 | 250 | 250 | 370 | 370 | ||
- *Power supply required is 3-phase 200/220/240/380/400/415/440V,50/60Hz. Voltage Down Transformer (Step-down transformer) is required for other voltage.
- Rated force of RANDOM follow the ISO5344 standard. Please contact us for further information on RANDOM force conditions of each system.
- Each value of specification indicates maximum power of the system. In the case of a long hours of test, it needs to be operated less than 70% of maximum power.
- In the case of RANDOM durability test, please set the test definition of the peak value of acceleration waveform to be operated less than 1400 m/s2.
- In the band of more than 2000Hz, excitation force pitches at the rate of -6dB/oct.
(The continuous use could be the cause of a premature failure.)
★ E/L is required for exporting.(See P.4)
- Specifications (High Power Water Cooled Type)
| System Model | EMK0301 | EMK0621 | EMK0622 | EMK0801 | EMK1251 | EMK1252 | EMK2001 | EMK2002 | ||
|---|---|---|---|---|---|---|---|---|---|---|
![]() |
Freq. Rang(Hz) | 0~3000 | 0~2500 | 0~2500 | 0~2500 | 0~2500 | 0~2500 | 0~2000 | 0~2000 | |
![]() |
SINE(kN) | 30.8 | 49.0 | 61.7 | 80 | 100 | 125 | 160 | 200 | |
| RANDOM(kNrms)2) | 21.5 | 49.0 | 61.7 | 80 | 1002) | 1252) | 1602) | 2002) | ||
| SHOCK(kN) | 61.6 | 98 | 123.4 | 160 | 200 | 250 | 320 | 400 | ||
|
SINE(m/s2) | 1000 | 1000 | 1000 | 1000 | 1000 | 1000 | 941 | 1000 | |
| RANDOM(m/s2rms) | 700 | 700 | 700 | 700 | 700 | 700 | 658 | 700 | ||
| SHOCK (m/s2) | 2000 | 2000 | 2000 | 2000 | 2000 | 2000 | 1882 | 2000 | ||
![]() |
SINE(m/s) | 1.85) | 2.45) | 2.45) | 2.45) | 2.45) | 2.45) | 1.85) | 1.85) | |
| SHOCK(m/s) | 1.8 | 2.0 | 2.0 | 2.0 | 2.0 | 2.0 | 2.0 | 2.0 | ||
![]() |
SINE(mmp-p) | 51 | 51 | 51 | 51 | 51 | 51 | 51 | 51 | |
| Max. TRAVEL(mmp-p) | 58 | 58 | 60 | 59 | 62 | 62 | 60 | 60 | ||
| Max. Load.(kg) | 500 | 1000 | 1000 | 1000 | 2000 | 2000 | 2000 | 2000 | ||
| Power Requirements(kVA)※1 | 49 | 76 | 87 | 100 | 150 | 170 | 200 | 280 | ||
|
Model | K030 | K060 | K060 | K080 | K125 | K125 | K200 | K200 | |
| Armature Mass(kg) | 27 | 40 | 40 | 60 | 70 | 70 | 170 | 170 | ||
| Armature Diameter(φmm) | 320 | 400 | 400 | 446 | 560 | 560 | 560 | 560 | ||
| Allowable eccentric moment(N・m) | 980 | 980 | 980 | 1550 | 2450 | 2450 | 4900 | 4900 | ||
| Dimensions(mm)W×H×D | 1100× 1090× 824 |
1380× 1085× 1000 |
1380× 1085× 1000 |
1595× 1050× 1200 |
1776× 1360× 1300 |
1776× 1360× 1300 |
2415× 1843× 1740 |
2415× 1843× 1740 |
||
| Mass(kg) | 3000 | 3700 | 3700 | 5000 | 7000 | 7000 | 13000 | 13000 | ||
![]() |
Model | SA4M- K30EM |
SA6M- K60EM |
SA8M- K60EM |
SA10M- K80EM |
SA13M- K125EM |
SA16M- K125EM |
SA23M- K200EM |
SA28M- K200EM |
|
| Max. Output(kVA) | 33 | 43 | 60 | 100 | 98 | 124 | 230 | 280 | ||
| Dimensions(mm)W×H×D | 1160 1950 824 |
1160 1950 824 |
1160 1950 824 |
1160 1950 824 |
1740 1950 850 |
1740 1950 850 |
2900 1950 850 |
2900 1950 850 |
||
| Mass(kg) | 1300 | 1350 | 1400 | 1750 | 2250 | 2350 | 3350 | 3500 | ||
![]() |
Vibration Controller | See Vibration Controller K2 | ||||||||
![]() |
Cooling Method | Shaker:Water Cooling/Amp:Air Cooling | ||||||||
| Primary Cooling Water Supply(l/min) | 195 | 260 | 260 | 390 | 390 | 390 | 650 | 650 | ||
![]() |
Dimensions(mm)W×H×D | 580 1700 850 |
580 1700 850 |
580 1700 850 |
580 1700 850 |
580 1700 850 |
580 1700 850 |
950 1900 800 |
950 1900 800 |
|
| Mass(kg) | 400 | 400 | 400 | 400 | 400 | 400 | 600 | 600 | ||
- *Power supply required is 3-phase 200/220/240/380/400/415/440V,50/60Hz. Voltage Down Transformer (Step-down transformer) is required for other voltage.
- Rated force of RANDOM follow the ISO5344 standard. Please contact us for further information on RANDOM force conditions of each system.
- Each value of specification indicates maximum power of the system. In the case of a long hours of test, it needs to be operated less than 70% of maximum power.
- In the case of RANDOM durability test, please set the test definition of the peak value of acceleration waveform to be operated less than 1400 m/s2.
- In SINE test, excitation time at the maximum velocity value in the sheet is limited within 1 minute. If the tests (SWEEP or SPOT) need high velocity
- It is system capability including an output transformer.
(The continuous use could be the cause of a premature failure.)
for more than 1 minute, the maximum velocity value should be reduced to 1.4m/s.
★ E/L is required for exporting.(See P.4)